The Metrological Traceability, Performance and Precision of European Radon Calibration Facilities - Archive ouverte HAL Access content directly
Journal Articles International Journal of Environmental Research and Public Health Year : 2021

The Metrological Traceability, Performance and Precision of European Radon Calibration Facilities

(1) , , (2) , , , , , , (3) , (4) , , , , (5) , , , , , , , , , (6) , , , (3) , , (7)
1
2
3
4
5
6
7
Andrei Antohe
  • Function : Author
  • PersonId : 1132497
Alexandra Cucoş
  • Function : Author
Eliska Fialova
Claudia Grossi
Kinga Hening
  • Function : Author
Jens Jensen
  • Function : Author
Patrick Lobner
  • Function : Author
Aurelian Luca
  • Function : Author
Franz Josef Maringer
Petr Otahal
  • Function : Author
Luis Quindós
  • Function : Author
Daniel Rábago
Carlos Sainz
László Szűcs
  • Function : Author
Constantin Teodorescu
  • Function : Author
Cathrin Tolinsson
  • Function : Author
Cornel Liviu Tugulan
  • Function : Author
Arturo Vargas
Josef Vosahlik
  • Function : Author
Hannah Wiedner
  • Function : Author

Abstract

An interlaboratory comparison for European radon calibration facilities was conducted to evaluate the establishment of a harmonized quality level for the activity concentration of radon in air and to demonstrate the performance of the facilities when calibrating measurement instruments for radon. Fifteen calibration facilities from 13 different European countries participated. They represented different levels in the metrological hierarchy: national metrology institutes and designated institutes, national authorities for radiation protection and participants from universities. The interlaboratory comparison was conducted by the German Federal Office for Radiation Protection (BfS) and took place from 2018 to 2020. Participants were requested to measure radon in atmospheres of their own facilities according to their own procedures and requirements for metrological traceability. A measurement device with suitable properties was used to determine the comparison values. The results of the comparison showed that the radon activity concentrations that were determined by European calibration facilities complying with metrological traceability requirements were consistent with each other and had common mean values. The deviations from these values were normally distributed. The range of variation of the common mean value was a measure of the degree of agreement between the participants. For exposures above 1000 Bq/m3, the variation was about 4% for a level of confidence of approximately 95% (k=2). For lower exposure levels, the variation increased to about 6%.
Fichier principal
Vignette du fichier
ijerph-18-12150.PDF (9.92 Mo) Télécharger le fichier
Origin : Publisher files allowed on an open archive

Dates and versions

irsn-03641836 , version 1 (14-04-2022)

Licence

Attribution - CC BY 4.0

Identifiers

Cite

Thomas Beck, Andrei Antohe, Francesco Cardellini, Alexandra Cucoş, Eliska Fialova, et al.. The Metrological Traceability, Performance and Precision of European Radon Calibration Facilities. International Journal of Environmental Research and Public Health, 2021, 18 (22), pp.12150. ⟨10.3390/ijerph182212150⟩. ⟨irsn-03641836⟩